Orbisphere M1100 Luminescent dissolved oxygen sensor, 0-2 ppm, with 28 mm Orbisphere fitting

Orbisphere M1100 Luminescent dissolved oxygen sensor, 0-2 ppm, with 28 mm Orbisphere fitting
Продукт №: M1100-S00
Единична цена Свържете се с нас
Доставка до 2 седмици

Minimal Drift and Annual Calibration

The M1100 sensor provides immediate oxygen readings with a measurement frequency of two seconds. The instrument carries consistent readings with no drift for 12 months without calibration (when utilising standard weekly CIP processes), surpassing other optical sensors that display significant drift after only a few months in similar conditions.

Highly accurate ppb DO measurement

The M1100 pioneered the use of luminescent technology in brewing applications (bypass) and as a result, the sensor doesn’t require the replacement of membranes or any electrolytes. Additionally, the sensor’s accuracy is unaffected by process changes or pressure shocks, further reducing maintenance. Annual maintenance is limited to just a few minutes for a zero-point calibration. Chemicals are not required for this process, making the task easier and safer without reducing measurement precision.

Robust optical technology without membrane or electrolyte

The M1100-L sensor has a lower detection level of 0.6 ppb. This highly accurate instrument’s readings are essential to control low oxygen levels in beverage production. As an example, it is critical to control low level oxygen in the post fermentation applications of the brewery to ensure high quality of the final product.

High Level Oxygen Measurement with Accurate ppm

The M1100-H sensor has a range of 0-40 ppm, and is ideal for use in wort applications, for example. Even in this harsh wort environment, the instrument maintains very good accuracy and minimal drift. Only a yearly maintenance and calibration will be required in most cases.

Какво има в кутията

Sensor only and manual.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.

Спецификации

IP клас на корпуса: Sensor resistant to all common CIP methods
Време за отговор : (90%)< 10 s (gas phase)
(90%)< 30 s (beer process)
Възпроизводимост: ± 0,8 ppb или 2 % в зависимост от това кое е по-голямо
Гаранция: 24 месеца
Дебит: 28 mm Orbisphere fitting
  50 - 300 mL/min
Дисплей, резолюция: 0.1 ppb
Измерване, диапазон: 0 - 2000 ppb DO (dissolved oxygen)
Какво има в кутията: Sensor only and manual.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.
Калибриране: Single point zero calibration with standard 99.999% nitrogen (quality 50) or equivalent oxygen free gas
Материал: 316 L Stainless Steel (1.4435), Silicone
Навлажнени материали: 316 L Stainless Steel (1.4435), Silicone
Най-долна граница на откриваемост: 0.6 ppb
Околна температура: -5 - 100 °C (23 - 212 °F)
Параметър: Oxygen
Повторяемост: ± 0.4 ppb или 1 % в зависимост от това, кое е по-голямо
Приложение: Beverage (bypass)
Проба: налягане: 1 - 20 bar абсолютна (14,5 до 290 psi)
Сертификати: 2004/108/EC - EN 61326-1
Тегло: M1100 12 mm: 0.6 kg
M1100 28 mm: 0.74 kg
Температурен диапазон: Accurate from -5 - 50 °C (23 - 122 °F)
Resistant - temperature from -5 - 100 °C (23 - 212 °F)
Точност: ± 0.8 ppb or 2 % whichever is greater
Условия на работа: 0 - 95 % non-condensing relative humidity