Demineralised water:
For rinsing
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Precision:
Better than 2% full scale range for standard test solutions
Reagent Requirements:
Keep between 10 - 30 °C
Validation:
Automatic; frequency freely programmable
Авариен сигнал:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Аналогови изходи:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Бр. потоци проби:
1, 2, 4 or 8
Гаранция:
24 месеца
Граница на откриваемост:
Calcium Hardness: ≤ 10 mg/L
Total Alkalinity: ≤ 20 mg/L
Дренаж:
Atmospheric pressure, vented, min. Ø 32 mm
Захранване:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Измерване, диапазон:
Ca: 10 - 100 mg/L CaCO₃; Alk: 20 - 500 mg/L CaCO₃
Calcium Hardness: 10 - 100 mg/L CaCO₃; Total Alkalinity: 20 - 500 mg/L CaCO₃
Измерване метод:
Acid-base titration (alkalinity only); Calmagite EDTA titration with LED dipping probe (combined)
Калибриране:
N.A.
Клас на защита:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Кювети за проба:
20 - 30 minutes
40 minutes (+ Total Hardness Ca)
Материал:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Метод, име:
Titrimetric
Околна температура:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Параметър:
Calcium Hardness; Total Alkalinity
Проба, дебит:
100 - 300 mL/min
Проба, качество:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Проба: налягане:
By external overflow vessel
Размери (В х Ш х Д):
690 мм x 465 мм x 330 мм
Сертификати:
CE compliant / UL certified
Смущения:
Hardness: Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA. Suspended or colloidal organic matter also may interfere with the end point. Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Alkalinity: Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
Тегло:
25 kg
Температура на проба:
10 - 30 °C