Power requirements (Amps):
2,5 A
Pressure Regulator Included?:
No
Авариен сигнал:
Two alarms selectable for sample concentration alarm, analyser system warning, or analyser system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Аксесоари:
Нищо
Влажност при работа:
Up to 90 % at 40 °C (104 °F) maximum
Вход:
¼" OD polyethylene tube, quick-disconnect fitting
Вход, налягане:
1 to 5 psi (0.07 - 0.34 bar), 1.5 psi (0.103 bar) is optimum
Възможности за комуникация:
4-20 mA output
Гаранция:
24 месеца
Граница на откриваемост:
0.03 mg/L (ppm)
Диапазон на налягането:
Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar)
Дисплей: течнокристален (фоново осветление):
LCD, 3.5 inch digit measurement readout and six-character alphanumeric scrolling text line
Изисквания към захранването (Hz):
50/60 Hz
Изисквания към захранването (напрежение):
100 - 115/230 V AC
Измерване, диапазон:
0 - 5 mg/L Cl₂
Какво има в кутията:
CL17 Chlorine Analyser with 1 Month Supply of Total Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Кювети за проба:
2.5 min
Монтаж, вид:
Wall mount
Обем на доставката:
CL17 Chlorine Analyser with 1 Month Supply of Total Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Ограждение:
ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Описание:
CL17 Chlorine Analyzer with Reagents for Total Chlorine
Параметър:
Chlorine, Free & Total
Повторяемост:
0.01 ppm or 3.0 % , whichever is greater
Проба, дебит:
200 - 500 mL per minute minimum required
Продухване с въздух:
0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psi (ca.1.4 bar) with ¼" OD tube, quick disconnect fitting
Работен температурен диапазон:
5 - 40 °C
Рекордер, изходи:
One 0/4-20 mA with an output span programmable over any portion of the 0 to 5 mg/L range. Recommended load impedance 3.6 to 500 ohms, 130 V isolation from earth ground.
Светлинен източник:
Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life
С включени аксесоари:
Нищо
Смущения:
Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference:
1 mg/L Cr6+ = approximately 0.02 mg/L as Cl2.
Hardness must not exceed 1,000 mg/L CaCO3.
Тегло:
11,3 kg
Температура на проба:
5 - 40 °C (41 - 104 °F)
Точност: