Demineralised water:
                                            
                                            
                                        
									    For rinsing / dilution
                                    
                                
    						
								
                                    
                                        
                                            
                                                Digital outputs:
                                            
                                            
                                        
									    Optional: Modbus (TCP/IP, RS485)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Earth connection:
                                            
                                            
                                        
									    Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
                                    
                                
    						
								
                                    
                                        
                                            
                                                Instrument air:
                                            
                                            
                                        
									    Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
                                    
                                
    						
								
                                    
                                        
                                            
                                                Precision:
                                            
                                            
                                        
									    Better than 2% full scale range for standard test solutions
                                    
                                
    						
								
                                    
                                        
                                            
                                                Reagent Requirements:
                                            
                                            
                                        
									    Keep between 10 - 30 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Validation:
                                            
                                            
                                        
									    Automatic; frequency freely programmable
                                    
                                
    						
								
                                    
                                        
                                            
                                                Авариен сигнал:
                                            
                                            
                                        
									    1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
                                    
                                
    						
								
                                    
                                        
                                            
                                                Аналогови изходи:
                                            
                                            
                                        
									    Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Бр. потоци проби:
                                            
                                            
                                        
									    1, 2, 4 or 8
                                    
                                
    						
								
                                    
                                        
                                            
                                                Гаранция:
                                            
                                            
                                        
									    24 месеца
                                    
                                
    						
								
                                    
                                        
                                            
                                                Граница на откриваемост:
                                            
                                            
                                        
									    ≤ 10 mg/L
                                    
                                
    						
								
                                    
                                        
                                            
                                                Дренаж:
                                            
                                            
                                        
									    Atmospheric pressure, vented, min. Ø 32 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Захранване:
                                            
                                            
                                        
									    100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
                                    
                                
    						
								
                                    
                                        
                                            
                                                Измерване, диапазон:
                                            
                                            
                                        
									    100 - 5000 mg/L CaCO3
Optional:
10 - 500 mg/L CaCO3
25 - 1250 mg/L CaCO3
50 - 2500 mg/L CaCO3
Internal dispenser dilution (max. factor 100)
                                    
                                
                                    
                                        
                                            
                                            
                                                 
                                            
                                        
									    100 - 5000 mg/L CaCO₃
                                    
                                
    						
								
                                    
                                        
                                            
                                                Измерване метод:
                                            
                                            
                                        
									    Acid-base titration (alkalinity only); Calmagite EDTA titration with LED dipping probe (combined)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Какво има в кутията:
                                            
                                            
                                        
									    EZ5001 Total / Free Alkalinity Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Bracket, 1 x pH Electrode, Routine S7 Connection and 1 x empty 10L Reagent Container with Fittings (HCl Solution)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Калибриране:
                                            
                                            
                                        
									    N.A.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Клас на защита:
                                            
                                            
                                        
									    Protection class: Analyser cabinet: IP44 / Panel PC: IP65
                                    
                                
    						
								
                                    
                                        
                                            
                                                Кювети за проба:
                                            
                                            
                                        
									    20 - 30 minutes
40 minutes (+ Total Hardness Ca)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Материал:
                                            
                                            
                                        
									    Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
                                    
                                
    						
								
                                    
                                        
                                            
                                                Метод, име:
                                            
                                            
                                        
									    Titrimetric
                                    
                                
    						
								
                                    
                                        
                                            
                                                Околна температура:
                                            
                                            
                                        
									    10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Параметър:
                                            
                                            
                                        
									    Alkalinity, total; Alkalinity, free
                                    
                                
    						
								
                                    
                                        
                                            
                                                Проба, дебит:
                                            
                                            
                                        
									    100 - 300 mL/min
                                    
                                
    						
								
                                    
                                        
                                            
                                                Проба, качество:
                                            
                                            
                                        
									    Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
                                    
                                
    						
								
                                    
                                        
                                            
                                                Проба: налягане:
                                            
                                            
                                        
									    By external overflow vessel
                                    
                                
    						
								
                                    
                                        
                                            
                                                Размери (В х Ш х Д):
                                            
                                            
                                        
									    690 мм x 465 мм x 330 мм
                                    
                                
    						
								
                                    
                                        
                                            
                                                Сертификати:
                                            
                                            
                                        
									    CE compliant / UL certified
                                    
                                
    						
								
                                    
                                        
                                            
                                                Смущения:
                                            
                                            
                                        
									    Hardness: Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA. Suspended or colloidal organic matter also may interfere with the end point. Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Alkalinity: Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Тегло:
                                            
                                            
                                        
									    25 kg
                                    
                                
    						
								
                                    
                                        
                                            
                                                Температура на проба:
                                            
                                            
                                        
									    10 - 30 °C